UV-Vis-Spectrometer
UV-Vis-Spectrometer | Jasco, V-650
Location: ECRC, Egerlandstraße 3, 91058 Erlangen, Room H00.14-1
The Jasco V-650 is a dual beam spectrophotometer with a deuterium (190 – 350 nm) and a halogen (330 – 900 nm) light source and a photomultiplier tube detector for accurate measurements of low concentration samples. The spectral bandwidth can be adjusted to > 0.1 nm. Low stray light slit settings provide excellent linearity of up to 4 absorbance units.
Dr.-Ing. Marcus Fischer
Instrument Specification:
- Wavelength range: 190 to 900 nm
- Wavelength accuracy: ±0.2 nm
- Maximum scanning speed: 12,000 nm/min
- Photometric range (Abs): -2 to 4 Abs
- Photometric system: dual beam
- Spectral bandwidth: 0.1 to 10.00 nm
- Samples: liquid and solid (thin films, solar, glass)
- Cuvette holder for 10 mm cuvettes
- Thermostatted Single Position Cell Holder (RT – 90°C) with magnetic stirrer
- Diffuse reflectance probe Praying Mantis™ (description see below)
in situ Reaction Chambers | Praying Mantis™ Harrick Scientific Products:
- 10 mm diameter diffuse reflection sample cup (0.25 ml volume)
- High Temperature Reaction Chamber:
- 133 µPa – 11.5 MPa (SiO2 window).
- Temperatures up to 910°C (under vacuum)
- Low temperature reaction chamber:
- 133 µPa – 11.5 MPa (SiO2 window)
- Temperatures down to -150°C – 600°C (under vacuum)
More information on request.