Analytical Centrifugation

Analytical Centrifuge | LUM, LUMiSizer LS611 (Incl. Sector Cells)

Location: LFG, Haberstraße 9a, 91058 Erlangen, Room 01.814

Dispersion analyzer for the determination of particle size distribution, droplet size distribution, separation velocity, Sauter diameter, particle magnetization, density distribution, Hansen solubility/dispersibility parameters, sedimentation, creaming, agglomeration, flocculation, coalescence, emulsification, dispersibility, shelf life (comparative & predictive) and direct accelerated stability analysis.

M. Sc. Moritz Moß

  • Rotor speed: 100 to 3,000 rpm
  • Rotor: up to 12 samples in parallel
  • Particle size: 10 nm to 1000 µm (depending on particle density)
  • Samples: concentrated dispersions and sediments
  • Sample volume: 0.05 to 2.0 mL
  • Concentration: 00015 Vol% – 90 Vol%
  • Density: up to 22 g/cm3
  • Viscosity: 0.8 to 108 mPas
  • Measurement time: 1 second to 99 hours
  • Detector: transmission detector (470 nm/865 nm)
  • Temperature control: -11°C to 60°C, +/- 1K
  • Measuring cells: various materials and optical path lengths
  • Conformity: ISO 4624, DIN EN 15870
  • Measurement mode: sedimentation velocity

https://www.lum-gmbh.com/produkte/lumisizer.html

Further information to follow